GE Healthcare
Revision 1
Job Card PHY A030 - Generator and Gantry Disconnection Reconnection
Location within Generator
Conditioner Unit
MDR
Generator Interface Board
400PL2
Mains Distribution Board
200PL1
Power Line Filter Box
Supply Command Board
200PL2
Anode Starter Board 300PL2
Auto Transformer
HV Tank
Notes:
* - Because there are two cables with the same 9-pin D-type connector originating from the same conduit,
you must pay special attention to ensure that you re-connect them to the correct place.
Chapter 6
Service Information and Procedures Class A 2385072-16-8EN
Connection
In
Out
RS-232
Gray cable with the 9-pin D-type connector,
labeled Cond. Cab. Cond. 2226611 *
XP4
Black power cable with the IEC type 320 female
end, labeled XJ4 Cond. Cab. 2225470
XP5
Black power cable with the IEC type 320 female
end, labeled XP5 Cond. Cab. 2386021
XJ4
Gray cable with the 9-pin D-type connector,
labeled XJ4 GENE 400 PL2 2226623 *
XJ10
Blue cable with the 25-pin D-type connector,
labeled XJ10 GENE 400 PL2 2226613
XJ6
XJ7
XJ9
XJ1
Blue cable with labeled PH2, originating from a
XJ3
Pink cable with labeled PH1, originating from a
Earth
Green/yellow earth cable, originating from a thick
Bolt above
XJ1
TP10
XJ2
Gray cable labeled G300-PL2 XP1 2381701 with
exposed shielding ending with three black cables
Orange block
Black cable, originating from white cable labeled
to BR1
Orange block
Black cable, originating from white cable labeled
to TH1
Earth symbol
Thin black cable with red plastic end, labeled HT+
N/A
Page no. 480
Cable description or label
Red coolant pipe (no label)
Red coolant pipe (no label)
Black optical fiber labeled
400 PL2 XJ6 2381702
Black optical fiber labeled
400 PL2 XJ7 2381704
Black optical fiber labeled
400 PL2 XJ9 2381706
thick white cable labeled 2381711
thick white cable labeled 2381711
white cable labeled 2381711
Black cable with blue plastic end,
originating from white cable labeled
Gene W125 - 2381710
Black cable with red plastic end,
originating from white cable labeled
Gene W125 - 2381710
to a black connector.
Gene W125 - 2381710
Gene W125 - 2381710
GND / 2381701
HV Cable
Senographe DS
C
D
R
N/
N/
A
A
JC-PHY-A-030.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.