GE Healthcare
Revision 1
Job Card DIAG A1000 - Gantry Diagnostic
FRU is the Gantry CPU, which you must change (see
page
1131).
FRUs:
2-44 JC DIAG A1406 COLL BLADE OVERCURRENT
Diagnostic for:
142/003 error code (CAN code is 0x5300)
142/004 error code (CAN code is 0x5300)
142/005 error code (CAN code is 0x5300)
142/006 error code (CAN code is 0x5300)
1. Check the cable connections, shielding, and grounding.
2. Reboot the system to check whether the error persists. If the error persists, open the collimator cover
and check whether there is interference between the blade gear and cover (identified by scratches
on the cover).
If there has been interference between the blade gear and cover change the Collimator Set (see
Card D/R A280 - Collimator Set on page
If there has not been interference between the blade gear and cover, turn the corresponding blade
gear manually in two directions.
If there is a mechanical hard point or the resistance in the two directions is noticeably different
then check whether there is mechanical disconnection (screws, rails). If there is a mechanical dis-
connection try to recover the connection. If the mechanical disconnection cannot be recovered,
change the Collimator Set (see
If there is no a mechanical hard point or no noticeable resistance in the two directions, power on
the system to check whether there is abnormal noise or vibration during blade movement. If there
is abnormal noise or vibration, change the Collimator Set (see
Set on page
1369).
FRUs:
2-45 JC DIAG A1407 COLL FILTER OVERCURRENT
Diagnostic for:
142/007 error code (CAN code is 0x5300)
1. Check the cable connections, shielding, and grounding.
2. Reboot the system to check whether the error persists. If the error persists, dismount the collimator
box to be able to see spectral filter. Turn the filter wheel manually, to check for mechanical interfer-
ence:
3. If there is mechanical interference, there is a mechanical assembly problem. Check whether the Fil-
ter Wheel interferes with tube tilt plate screw. If the Filter Wheel interferes with tube tilt plate screw,
reassemble spectral filter and tilt plate.
If the interference is within filter itself, change the Filter Wheel (see
tor Filter Wheel Assembly on page
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Collimator Set
Filter Wheel
Gantry CPU
1369).
Job Card D/R A280 - Collimator Set on page
Collimator Set
1375).
Page no. 970
Job Card D/R A190 - Gantry CPU Board on
Job Card D/R A280 - Collimator
Job Card D/R A285 - Collima-
Senographe DS
Job
1369).
JC-DIAG-A-1000.fm