GE Senographe DS Manual page 919

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Code
Category
162/009
HALT
162/010
INHIBIT
162/011
INHIBIT
162/012
WARNING
162/013
HALT
162/014
HALT
162/015 ADVISORY LCD: A19 Stereotac-
162/016 ADVISORY LCD: A12 Move tube
162/017 ADVISORY
ERR-SUB-A-003.fm
Text in Log File or
on LCD
STEREOTIX HARD-
WARE DETECTION
IN FAILURE
STEREOTIX CAN
NOT ENTER
READY STATE
STEREOTIX CAN
NOT ENTER PUNC-
TURE STATE
STEREOTIX TAR-
GET LIST IS FULL
STEREOTIX NODE
NOT RESPONDING
INTERNAL SOFT-
WARE ERROR
tic Positioner is not
ready
head to center posi-
tion
LCD: A16 Rotation
limit for Stereotaxy is
reached
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
The Gantry CPU failed to detect the stereotix hard-
ware.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
The stereotix node can not switch to the READY
state.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
The stereotix node can not switch to the PUNCTURE
state.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
The stereotix node's target list is FULL.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
Heartbeat signal was not sent by the stereotix node.
The stereotix node sends a Heartbeat signal every
second to Gantry CPU. If not, the Gantry CPU cre-
ates this error. This error occurs when an operator
did not disconnect the stereotix within 10s after
releasing the latches (time out).
Operator error. After releasing the latches, operator
has 10s to disconnect the stereotix. Reboot the sys-
tem from the console then redo the operation into the
allocated timestamp.
The Gantry CPU received an unknown error code
from the stereotix node.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
The stereotix positioner is not ready.
Operator error. No service recommended actions.
Move to angulation 0°
Operator error. No service recommended actions.
Stereotactic rotation limit reached.
Operator error. No service recommended actions.
Page no. 919
ERR SUB A003 - Gantry Error Messages
Senographe DS
Job Card
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF