GE Senographe DS Manual page 574

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Job Card ELE A013 - AWS Configuration
2. To change any of the displayed items, type the new information and confirm the entry by clicking the
Submit button.
Note:
1. This screen supports all US ASCII characters (i.e. roman alphabet letters, numbers, and classic
symbols). All accented characters are not supported, and must be substituted by appropiate let-
ters.
2. The entry Station Name is the name (hostname) that is used for the system when interacting
with the RIS server. The Station Name must be the same value as the hostname that is defined in
the TCP/IP networking configuration of
uration on page
Do not use the following characters in the Station Name:
!
ONLY USE ALPHANUMERIC CHARACTERS WITHOUT SPACES.
It is also possible to use the characters - (dash) and . (dot).
Ensure that the Station Name is no more than 12 characters long.
3. If there is more than one Senographe system on the hospital site, the names chosen must clear-
ly differentiate between them. For example, by including the unit number or name).
6-3
Set General Configuration Parameters
1. Click General under Configuration to display the General Configuration screen.
2. To change any of the displayed items, type the new information and/or select a different option in the
following:
a. The System Clock yyyy mm dd and hh:mm:ss fields set the current date and time, respec-
tively.
! Notice:
After changing the clock the system must be shut down and restarted. If this is not done, system
behavior can be unpredictable, because several interacting processes use the clock.
Chapter 7
625.
$ "
'
'
{ }
[
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A021 - Networking Connection and Config-
]
*
@ # ? ^ _ / etc.
Page no. 574
Senographe DS
JC-ELE-A-013.fm

Hide quick links:

Advertisement

Table of Contents
loading