Job Card TSG A015 - Lift Movement Troubleshooting
Repeat these sub-steps above several times until the Lift Screw is cleaned. A well cleaned Lift Screw
looks like that shown below.
2. Apply Grease (for Lift and Rotation) (5212483) (Orapi CT FLON 607) to the Lift Screw by hand.
Ensure the Lift Screw is homogeneously covered with grease, as shown in illustration 8.
CAUTION
Do not use any other type of grease, other than Grease for Lift and Rotation (5212483)
(Orapi CT FLON 607).
Do not use a brush to apply the grease as this may introduce foreign objects on to the Lift
Screw. Only use your hands to apply the grease to the Lift Screw.
About 35 g (1.23 oz) of grease is sufficient to cover the entire length of the Lift Screw.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN