2-23 JC DIAG A1117 ROTATION COMMUNICATION FAILURE
Diagnostic for:
112/039 error code (CAN code is 0x8210) invalid PDO length
Change the Rotation Board
If the error persists, change the PL 103 CPU Board (see
page
1131).
If changing the CPU Board solved the error, put back the original Rotation Board
A221 - Rotation Board on page
FRUs:
2-24 JC DIAG A1118 ROTATION CONTROL FAILURE
Diagnostic for:
112/045 error code (CAN code is 0x5028) optical fork reached error
Inspect the optical fork cable. If the optical fork cable seems to be good, change the optical fork (see
Job Card D/R A223 - Rotation Optical Fork on page
If the error persists, put back the original optical fork and change the Rotation Board
A221 - Rotation Board on page
Reboot the Gantry. If the error is still present, put back the original Rotation Board as it is good
Card D/R A221 - Rotation Board on page
A190 - Gantry CPU Board on page
FRUs:
JC-DIAG-A-1000.fm
Service Information and Procedures Class A 2385072-16-8EN
(Job Card D/R A221 - Rotation Board on page
1223).
Rotation Board
Nephtys Gantry CPU
1223).
1223). Then change the CPU Board (see
1131).
Rotation Board
Nephtys Gantry CPU
Optical fork + cable
Page no. 955
Job Card DIAG A1000 - Gantry Diagnostic
Job Card D/R A190 - Gantry CPU Board on
1231).
Senographe DS
1223).
(Job Card D/R
(Job Card D/R
(Job
Job Card D/R
Chapter 9