GE Senographe DS Manual page 491

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
19. Release the two allen screws (18) (4 mm allen wrench) that secure the Generator cable bracket (19)
to the Generator framework. Remove the Generator cable bracket from the Generator.
Keep the two allen screws in a safe place as you will use them again later on to re-connect the Gen-
erator Cable bracket to the Generator framework.
JC-PHY-A-030.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A030 - Generator and Gantry Disconnection Reconnection
18
Page no. 491
19
Senographe DS
Chapter 6

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents