ERR SUB A007 - IDC Error Messages

1
HOW TO VIEW THE ERROR LOG
Errors reported by the IDC diagnostic routines are recorded in a common system error log. The common
system error log also contains error messages generated by the AWS and Gantry/Generator applica-
tions.
To view the error log content, right click the AWS Screen background; select Service Tools/Service
Desktop/Error Log from the drop-down menu.
The most recent error code generated by the IDC can also be viewed on the display on the rear of the
IDC computer.
2
IDC DIAGNOSTICS
Each time that the IDC sub-system is booted (powered up or reset), it carries out a series of diagnostic
self-tests, first under the control of the BIOS, then under software control. As each test is performed, a
corresponding code appears on the four-digit green LCD panel at the rear of the IDC Computer.
2-1
BIOS Diagnostic Display
While the BIOS tests are running, the LCD displays two blank digits followed a two-digit code corre-
sponding to the current test. If a test detects an error or fails to complete, the tests are halted and the
code of the failing test remains displayed.
2-2
Software Diagnostic Display
Three sets of software diagnostics are carried out. While these tests are running, the display shows
X000, where X may be 1, 2, or 3.
If all tests are completed without failure, the display shows FFFF.
If a test fails, the display shows a code in the form XXFF for two seconds. XX is a code correspond-
ing to the failing test. The tests then continue.
On completion of the tests after a failure occurs, the display shows F0FF.
3
ERROR DETECTION
If the IDC halts with a display in the format FFFF, all power-up tests have passed without failure.
If the IDC halts with a display in the format --XX, where XX can be any number, a BIOS test has
failed. If the problem occurs again after reset, the recommended action is to change the IDC Com-
puter.
If the IDC halts with a display in the format F0FF, a software test has failed. Reset the IDC and watch
the LCD display.
If the failure is repeated, a code XXFF appears for two seconds. Note the error code XX.
Table 1 gives a list of IDC diagnostic error codes, with recommended sequences of actions.
ERR-SUB-A-007.fm
Service Information and Procedures Class A 2385072-16-8EN
Page no. 937
Senographe DS
ERR SUB A007 - IDC Error Messages
Chapter 9
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF