AWS Power Checks
It is likely that the ADS computer PSU is faulty. Plug the ADS computer into a wall
socket. If the ADS computer still does not power up and function correctly then the
JC-TSG-A-001.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A001 - AWS Troubleshooting - No Display on Power Up
From
page 1010
UPS and MDR check is good, but
AWS does not power up,
Is the UPS-AWS
power cable
incorrectly connected
or defective?
No
Is the MDR-UPS
power cable
incorrectly connected
or defective?
No
ADS computer PSU is faulty.
Is the ADS power
supply faulty?
No
If the problem persists, start
again on
page
1010, or ask for
help from your OLC.
Page no. 1011
Correct the cable problem.
Yes
Correct the cable problem.
Yes
Change the ADS Computer
Yes
Senographe DS
Chapter 9