GE Healthcare
Revision 1
Job Card TSG A015 - Lift Movement Troubleshooting
ILLUSTRATION 9 - EXAMPLES OF WELL APPLIED GREASE AFTER LIFT MOVEMENT
If there are some areas where no grease exists, apply grease to these areas, move the Arm again in
slow and fast speeds, and check the greasing again.
3-4
Lift Gas Spring Checks
1. Remove the Tube Head covers, Bucky and Compression Paddle.
Note:
If you do not remove the Tube Head covers, Bucky and Compression Paddle, the Arm assembly
will be too heavy for the Lift Gas Spring checks to work.
2. Remove the two springs of curtains from the back of the Gantry. (Using a crook or tiewrap could
help). If necessary, move the trolley position so that the springs (1) that attach upper and lower cur-
tains are visible.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
1
Page no. 1052
Senographe DS
JC-TSG-A-015.fm