Job Card Diag A1000 - Gantry Diagnostic - GE Senographe DS Manual

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1

Job Card DIAG A1000 - Gantry Diagnostic

The Gantry software sends a warning message each time an abnormal situation is detected.
This Job Card provides a guide for when the warning message content is not self-sufficient. It is espe-
cially useful when the warning message indicates more than one defective FRU.
This Job Card gives a set of individual diagnostic procedures (identified as Job Cards), each corre-
sponding to one or several warning codes, as summarized below.
1
GANTRY DIAGNOSTIC JOB CARD SUMMARY
JC Number
JC DIAG A 1001
JC DIAG A 1004
JC DIAG A 1005
JC DIAG A 1008
JC DIAG A 1011
JC DIAG A 1012
JC DIAG A 1013
JC DIAG A 1014
JC DIAG A 1015
JC DIAG A 1016
JC DIAG A 1017
JC DIAG A 1103
JC DIAG A 1104
JC DIAG A 1107
JC DIAG A 1109
JC DIAG A 1110
JC DIAG A 1111
JC DIAG A 1112
JC DIAG A 1113
JC DIAG A 1114
JC DIAG A 1115
JC DIAG A 1116
JC DIAG A 1117
JC DIAG A 1118
JC DIAG A 1201
JC DIAG A 1203
JC DIAG A 1205
JC DIAG A 1206
JC-DIAG-A-1000.fm
Service Information and Procedures Class A 2385072-16-8EN
FOOTSWITCH IS PRESSED
LIFT CONTROL FAILURE
LIFT CONTROL FAILURE
LIFT CONTROL FAILURE
LIFT CONTROL FAILURE
LIFT CONTROL FAILURE
LIFT CONTROL FAILURE
LIFT POSITION OVERSHOOT
LIFT CONTROL FAILURE
LIFT CONTROL FAILURE
LIFT CONTROL FAILURE
ROTATION CONTROL FAILURE
ROTATION CONTROL FAILURE
ROTATION CONTROL FAILURE
ROTATION CONTROL FAILURE
ROTATION CONTROL FAILURE
ROTATION CLUTCH FAILURE
ROTATION BRAKE DEFECT
ROTATION CONTROL FAILURE
ROTATION CONTROL FAILURE
ROTATION POSITION OVERSHOOT/ROTATION
CONTROL FAILURE
ROTATION CONTROL FAILURE
ROTATION COMMUNICATION FAILURE
ROTATION CONTROL FAILURE
CONTROL PAD FAILURE
COMPRESSION CONTROL FAILURE
COMPRESSION CONTROL FAILURE
COMPRESSION CONTROL FAILURE
Page no. 943
Job Card DIAG A1000 - Gantry Diagnostic
Name
Senographe DS
Chapter 9
Page
page 945
page 946
page 946
page 947
page 947
page 948
page 948
page 948
page 949
page 949
page 950
page 950
page 950
page 951
page 951
page 952
page 952
page 953
page 953
page 953
page 954
page 954
page 955
page 955
page 956
page 957
page 957
page 958
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.