Job Card DIAG A1000 - Gantry Diagnostic
2-21 JC DIAG A1115 ROTATION POSITION OVERSHOOT/ROTATION CONTROL
FAILURE
Diagnostic for:
112/029 error code (CAN code is 0x6006) position error
112/030 error code (CAN code is 0x6007) deceleration error
Change the Rotation Board
If the error persists, change the rotation gas spring (see
page
1251).
FRUs:
2-22 JC DIAG A1116 ROTATION CONTROL FAILURE
Diagnostic for:
111/004 error code (CAN code is 0x9002) valid motion error
Inspect the cables listed below.
If you find any damage in a cable or connector, change the cable or the Rotation Board
R A221 - Rotation Board on page
order:
-
W213 - POS-BUS 1 cable for Lift to Rotation
-
W304A cable for paddle detector
-
W304B cable for paddle detector
-
W314 Sys Pos Bus cable
-
W311 - POS-BUS 2 cable for Bucky board
If the error persists, change the Rotation Board
If the error persists, change the PL 103 CPU Board (see
page
1131).
FRUs:
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
(Job Card D/R A221 - Rotation Board on page
Rotation Board
Gas spring for rotation comp
1223). If no defect can be seen, change the cables in the following
Motor + encoder with cable
Rotation Board
W314 Sys Pos Bus cable
W304A cable for paddle detector
W304B cable for paddle detector
W311 - POS-BUS 2 cable for Bucky board
W213 - POS-BUS 1 cable for Lift to Rotation
Nephtys Gantry CPU
Page no. 954
Job Card D/R A230 - Rotation Gas Spring on
(Job Card D/R A221 - Rotation Board on page
Job Card D/R A190 - Gantry CPU Board on
Senographe DS
1223).
(Job Card D/
1223).
JC-DIAG-A-1000.fm