GE Healthcare
Revision 1
ILLUSTRATION 3 - STEPS TO PERFORM IF BAD PIXELS EXIST AT DETECTOR EDGES
Perform post-corrective actions
(see
section 4-1
,
Calibration to
be performed after bad pixel
calibration on page
1026).
Perform post-corrective actions (see
section 4-1
,
Calibration to be per-
formed after bad pixel calibration on
page
1026).
1.
Change the Collimator
2.
Perform collimator calibration
3.
Perform Bad Pixel Calibration
Perform
post-corrective
(see
section 4-1
,
Calibration to be
performed after bad pixel calibration
on page
1026).
JC-TSG-A-010.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
(see
Job Card CAL A029 - Collimator Format Calibration
(see
Job Card CAL A041 - Bad Pixel Calibration on page
YES
Bad Pixel calibration passes?
1.
Adjust the Collimator positioning.
2.
Perform collimator calibration
tion
)
on page 1683
3.
Perform Bad Pixel Calibration
)
page 1693
YES
Bad Pixel calibration passes?
(see
Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
(see
Job Card CAL A029 - Collimator Format Calibration
(see
Job Card CAL A041 - Bad Pixel Calibration
actions
YES
Bad Pixel calibration passes?
The customer must not use the system.
Use the service escalation process.
Perform collimator calibration
Perform Bad Pixel Calibration
NO
(see
Job Card CAL A029 - Collimator Format Calibra-
(see
Job Card CAL A041 - Bad Pixel Calibration
NO
NO
Page no. 1025
Senographe DS
)
on page 1683
1693)
on
on page 1019
)
on page 1683
)
on page 1693
)
Chapter 9
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated