GE Healthcare
Revision 1
Job Card D/R A150 - System Fuses
6
PROCEDURE
1. Refer to the appropriate illustration for the position of the fuse you want to change.
•
For the Main Distribution Assembly fuse locations, refer to illustration 1.
•
For the Generator Command Board fuse locations, refer to illustration 2.
•
For the Generator Supply Command Board fuse locations, refer to illustration 3.
•
For the Generator Main fuse locations, refer to illustration 4.
•
For the Mains Distribution Board fuse locations, refer to illustration 5.
•
For the PDU Board fuse locations, refer to illustration 6.
•
For the MPCB Board fuse locations, refer to illustration 6.
2. If changing a fuse within the Generator, ensure that the components in the Generator are fully dis-
charged according to the steps described in
Discharged on page
3. Change the fuse.
4. Power up the system and ensure that the system operates as expected without error.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
1110.
section
5-1,
Ensure Capacitors on Generator Boards are
Page no. 1118
Senographe DS
JC-DR-A-150.fm
Need help?