GE Healthcare
Revision 1
2-4
Power-Up Sequence – B
200PL2 DE1 and DE2
blinking with console
button and 200PL1 DE1
Recent repeated
power–up attempts?
Reset power–up sequence by
opening Generator mains supply
switch S1 ("0" position), then
closing it again ("1" position).
Normal power up
sequence now possible
using console
JC-TSG-A-020.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A020 - Generator Power-Up Sequence Flow Charts
B
No
lit?
Yes
No
Yes
button?
No
Yes
End.
300PL1 600VF,
600VR, 0/600VF, 0/
600VR connections
200PL1 lugs W1,
W2, W3, and 300VR
connections tight?
300PL1: Voltage
between 600VR and
0/600VR at approx.
200PL1: Voltage
between PT13 and
PT18 at approx.
Change 200PL2.
No
tight?
Yes
No
Yes
No
600 V DC?
Yes
No
3 V DC?
Yes
Page no. 1073
Senographe DS
Tighten appropriate
connections.
Tighten appropriate
connections.
Change capacitor
bank.
Change 200PL1.
Chapter 9
Need help?