Revision 1
Job Card D/R A141 - Image Receptor/Digital Detector
4. Insert the SFOV Mammo Detector Manufacturing Data CD-ROM into the CD-ROM drive of the AWS
unit.
5. Click the START button to start the digital detector firmware update process.
The Results section of the Service Desktop updates with a Processing progress bar while the firm-
ware is being written to the digital detector.
Once the firmware is written to the digital detector the Results section of the Service Desktop
updates with the message DETECTOR REPLACEMENT COMPLETE, and the AWS unit automati-
cally ejects the CD-ROM drive.
6. Remove the SFOV Mammo Detector Manufacturing Data CD-ROM from the CD-ROM drive of the
AWS unit, and close the CD-ROM drive.
7. Reboot the Senographe system.
8. View the Service Desktop home page (see
the value of the DETECTOR_ID on the Service Desktop home page corresponds to the value written
on the SFOV Mammo Detector Manufacturing Data CD-ROM. The DETECTOR_ID value is of the
format LLnnn_nn (e.g. FL002_01).
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Launching the Service Desktop on page
Page no. 1104
Senographe DS
120). Check that
JC-DR-A-141.fm