Job Card DIAG A005 - Error Log Retrieval
9. Ensure that the ADS is reachable by entering the following ping command:
ping 192.168.1.10
The Command Window should update similar to the following.
10. Ensure that the IDC is reachable by entering the following ping command:
ping 192.168.1.20
The Command Window should update similar to the following.
5-2-3
Getting ADS Log Files
The following steps describe getting the ADS log files on to your laptop via FTP, using the Command
Window. Alternatively, you can use a graphical FTP client such as HummingBird FTP Client, CoreFTP or
FileZilla to get the ADS log files. The principles of getting the ADS log files are the same as those below
(i.e. IP address, username/password, file locations are all the same, and transfer mode is Binary (not
ASCII)).
1. On your laptop, open a Command Window (Start > Run, then enter cmd and click OK).
The path of the Command Window will be D:\Documents and Settings\<your SSO>.
2. Create a new directory within D:\Documents and Settings\<your SSO> with the date format
YYMMDD, by entering:
mkdir YYMMDD
where YYMMDD represents the current date (e.g. 081117 for 17th November 2008).
3. Change into the YYMMDD directory, by entering:
cd YYMMDD