GE Senographe DS Manual page 674

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Job Card ELE A023 - Declare Worklist Hosts
6. Click Query. The Query definition window like the following appears.
7. Select All systems in modality from the Which system section.
! Notice:
If you do not get a response when "refreshing" using All systems in modality, select All
systems.
Worklist servers cannot always handle a filtered query such as "All systems in modality". This is
since the query places the AWS hostname in the DICOM tag (0040,0001)= "Scheduled Station AE
Title".
8. Click Save as default to freeze the selection.
7
TROUBLESHOOTING
Use the "Ping" command to test access to the worklist server.
Use the "SendEcho" command to test DICOM association (most worklist servers can support it).
Double check AE Title declaration on each side.
See log ~sdc/senovision/logfiles/trace.Worklist.log.
Chapter 7
Service Information and Procedures Class A 2385072-16-8EN
Page no. 674
Senographe DS
JC-ELE-A-023.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Table of Contents

Save PDF