GE Senographe DS Manual page 921

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Code
Category
163/000
INHIBIT
163/001
INHIBIT
163/005
HALT
163/006
HALT
163/007
HALT
163/105
WARNING
164/001
INHIBIT
164/002
INHIBIT
164/003
INHIBIT
ERR-SUB-A-003.fm
Service Information and Procedures Class A 2385072-16-8EN
Text in Log File or
on LCD
INVALID BUCKY
TYPE
INVALID PADDLE
DETECTED
INTERNAL SOFT-
WARE ERROR
INTERNAL SOFT-
WARE ERROR
INTERNAL SOFT-
WARE ERROR
INVALID LAN-
GUAGE. DEFAULT
USED
BLADE FORMAT
OUT OF FOV
INTERNAL SOFT-
WARE ERROR
COMMUNICATION
ERROR WITH COL-
LIMATOR NODE
Possible Cause/Recommended Action
The Gantry CPU received an invalid image receptor
type.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry CPU received an invalid paddle type.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry CPU received an invalid button function
identity.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry CPU received an invalid button state.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
There is a button state inconsistency. The lift or rota-
tion button counter on the Gantry CPU has overflown
or underflown.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
An invalid language is currently configured on, and
English is currently being used.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The blade configuration in the collimator is outside
the current field of view (FOV) setting.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry CPU received a machine event with an
unexpected state.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
A timeout command on the Gantry CPU is waiting for
a response from the collimator.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Page no. 921
ERR SUB A003 - Gantry Error Messages
Senographe DS
Job Card
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF