Revision 1
Cable Lay-out and Pin-out
1-17. W114/W116/W209 - PDU Board J7 Connections
F u se  2  o n  
T e rm in a l B lo c k  
S T 1 0 2
In te rfa c e
B o a rd
P L 1 0 4
L ift
B o a rd
P L 2 0 1
W114 - Column Fuse Check
PDU Board PL101 J7 Pin 1 to Terminal 2 ST102-F1/1.
W116 - PL104 power
PDU Board PL101 J7 to Interface Board PL104 J1.
1
W209 - 48 V LP
PDU board PL101 J7 to Lift Board PL201 J2:
1-18. W116 - PL104 Power
See
W114/W116/W209 - PDU Board J7 Connections on page
1-19. W117 - Terminal Block Interconnection
Terminal block ST102 from F1/2 to F2/2.
1-20. W119 - Serial Link TX Generator to Interface Board
Generator Interface J7 to Interface Board PL104 J9.
Chapter 3
J1
J2
Pin no.
PL 104 J1-1
PL 104 J1-2
Pin no.
PL 201 J2-1
PL 201 J2-2
Service Information and Procedures Class A 2385072-16-8EN
W 1 1 4
W 1 1 6
W 2 0 9
Level
+48 V
GND
Functions
+ 48 Volts (max. 100 mA)
GND
Page no. 302
J7
1 0
1 1
1 2
Pin no.
PL 101 J7-5
PL 101 J7-11
Pin no.
PL 101 J7-3
PL 101 J7-9
302.
Senographe DS
P D U
B o a rd