Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
4-5
Calibration to be performed after tube tilt calibration
Calibration/Test
Tube Tilt Calibration
Collimator Format Calibration
Bad Pixel Calibration
Gain calibration
Flat field tests
Conversion Factor
Half Value Layer
AOP Calibration
mAs non linearity
AOP/SNR Check
If AOP/SNR test is not OK, repeat all calibrations from AOP Calibration
4-6
Calibration to be performed after changing or moving the collimator
Calibration/Test
Collimator Format Calibra-
tion
Collimation Checks
Bad Pixel Calibration
Gain calibration
Flat field tests
Conversion Factor
Half Value Layer
AOP Calibration
mAs non linearity
If AOP/SNR test is not OK, repeat all calibrations from AOP Calibration
4-7
Calibration to be performed after collimator format calibration
Calibration/Test
Collimation Checks
Bad Pixel Calibration
Gain calibration
Conversion Factor
Half Value Layer
AOP Calibration
mAs non linearity
If AOP/SNR test is not OK, repeat all calibrations from AOP Calibration
Chapter 9
Job Card CAL A027 - Tube Tilt Calibration on page 1681
Job Card CAL A029 - Collimator Format Calibration on page 1683
Job Card CAL A041 - Bad Pixel Calibration on page 1693
Job Card CAL A043 - Detector Gain Calibration on page 1697
Job Card ELE A032 – IQ Tools Flat Field Test on page 699
Job Card CAL A045 - Conversion Factor Measurement on page 1701
Job Card CAL A046 - Half-Value Layer Measurement on page 1703
Job Card CAL A053 - mAs Non-linearity Calibration on page 1725
Job Card ELE A038 - AOP Mode and SNR Check on page 707
Job Card CAL A029 - Collimator Format Calibration on page 1683
Job Card ELE A028 - Collimation Checks on page 679
If Collimator Check is not OK, repeat Collimator Calibration
Job Card CAL A041 - Bad Pixel Calibration on page 1693
Job Card CAL A043 - Detector Gain Calibration on page 1697
Job Card ELE A032 – IQ Tools Flat Field Test on page 699
Job Card CAL A045 - Conversion Factor Measurement on page 1701
Job Card CAL A046 - Half-Value Layer Measurement on page 1703
Job Card CAL A051 - AOP Calibration on page 1723
Job Card CAL A053 - mAs Non-linearity Calibration on page 1725
If Collimator Check is not OK, repeat Collimator Calibration
Job Card CAL A043 - Detector Gain Calibration on page 1697
Job Card CAL A045 - Conversion Factor Measurement on page 1701
Job Card CAL A046 - Half-Value Layer Measurement on page 1703
Job Card CAL A053 - mAs Non-linearity Calibration on page 1725
Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A051 - AOP Calibration on page 1723
Job Card
Job Card ELE A028 - Collimation Checks on page 679
Job Card CAL A041 - Bad Pixel Calibration on page 1693
Job Card CAL A051 - AOP Calibration on page 1723
Page no. 1028
Job Card
Job Card
Senographe DS
JC-TSG-A-010.fm