GE Healthcare
Revision 1
Central Listing
The on/off power buttons on the X-ray Console override the power states of each component. For exam-
ple, if you power off/on the system from the X-ray Console, the states of the power to each of the compo-
nent defaults as indicated below.
Push button
Conditioner
Detector Power
Supply
Cart
The table above assumes that the mains AC power supply is connected to the Generator, and the Gen-
erator/MDR breakers are switched on.
Chapter 3
Component controlled
Conditioner
Detector Power Supply
UPS in Control Station
(supplies power to entire
Control Station)
Service Information and Procedures Class A 2385072-16-8EN
Default status when X-ray
Console ON button sup-
plies power to the system
ON
ON
ON
Page no. 280
Senographe DS
Default status when X-ray
Console OFF button cuts
power to the system
ON
ON
OFF
Indicators and Switches.fm