Revision 1
ERR SUB A003 - Gantry Error Messages
Code
Category
112/059
DEBUG
ONLY
112/060
DEBUG
ONLY
112/061
DEBUG
ONLY
112/062
DEBUG
ONLY
112/063
DEBUG
ONLY
112/064
HALT
112/065
HALT
112/066
DEBUG
ONLY
112/067
DEBUG
ONLY
112/068
DEBUG
ONLY
112/069
DEBUG
ONLY
112/070
DEBUG
ONLY
112/071
DEBUG
ONLY
Chapter 9
Text in Log File or
on LCD
ROTATION INVALID
INITIALIZATION
ERROR
ROTATION INVALID
STEP LOST ERROR
ROTATION INVALID
FORCE SENSOR
DEFECT
ROTATION INVALID
COMMAND ERROR
ROTATION INVALID
FORCE SENSOR
RANGE ERROR
ROTATION INVALID
SW ERROR
ROTATION INVALID
INTERNAL SW
ERROR
ROTATION INVALID
LIFT DOWN PEDAL
PRESS ERROR
ROTATION INVALID
LIFT UP PEDAL
PRESS ERROR
ROTATION INVALID
COMPRESSION
PEDAL PRESS
ERROR
ROTATION INVALID
DECOMPRESSION
PEDAL PRESS
ERROR
ROTATION INVALID
LIFT UP SLOW
BUTTON PRESS
ERROR
ROTATION INVALID
LIFT UP FAST BUT-
TON PRESS
ERROR
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The lift down pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The lift up pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The compression pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The Lift Up button on either the Compression Arm or
the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The Lift Up button on either the Compression Arm or
the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The Down Lift button on either the Compression Arm
or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.