ERR SUB A003 - Gantry Error Messages
Code
Category
112/059
DEBUG
ONLY
112/060
DEBUG
ONLY
112/061
DEBUG
ONLY
112/062
DEBUG
ONLY
112/063
DEBUG
ONLY
112/064
HALT
112/065
HALT
112/066
DEBUG
ONLY
112/067
DEBUG
ONLY
112/068
DEBUG
ONLY
112/069
DEBUG
ONLY
112/070
DEBUG
ONLY
112/071
DEBUG
ONLY
Chapter 9
Text in Log File or
on LCD
ROTATION INVALID
INITIALIZATION
ERROR
ROTATION INVALID
STEP LOST ERROR
ROTATION INVALID
FORCE SENSOR
DEFECT
ROTATION INVALID
COMMAND ERROR
ROTATION INVALID
FORCE SENSOR
RANGE ERROR
ROTATION INVALID
SW ERROR
ROTATION INVALID
INTERNAL SW
ERROR
ROTATION INVALID
LIFT DOWN PEDAL
PRESS ERROR
ROTATION INVALID
LIFT UP PEDAL
PRESS ERROR
ROTATION INVALID
COMPRESSION
PEDAL PRESS
ERROR
ROTATION INVALID
DECOMPRESSION
PEDAL PRESS
ERROR
ROTATION INVALID
LIFT UP SLOW
BUTTON PRESS
ERROR
ROTATION INVALID
LIFT UP FAST BUT-
TON PRESS
ERROR
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The lift down pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The lift up pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The compression pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The Lift Up button on either the Compression Arm or
the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The Lift Up button on either the Compression Arm or
the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
The Down Lift button on either the Compression Arm
or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
Page no. 876
Senographe DS
Job Card
ERR-SUB-A-003.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.