Potentiometer Upper Reference Voltage.
The Gantry LCD updates with the message "Lift node fully calibrated. Calibration can be saved or
canceled on ADS." and the AWS screen updates instructing you to either save or cancel the calibra-
tion (see
Selection of the Lift Travel Limit (Completion) on page
! Notice:
If you do not move the lift up to its highest possible position until it automatically stops, and you
press the FOV size button the Gantry LCD updates with the message "Upper lift position out of
range. Check lift position or positioner signal! Then press FOV button.". In this case simply contin-
ue to move the lift up to its highest possible position until it automatically stops and then press the
FOV size button.
6-2-4
Selection of the Lift Travel Limit (Completion)
Use the following steps to save the calibration results and set the lift travel limit,:
1. On the Lift Position and Upper Limit Calibration page, click the Save button.
A Lift Optical Fork Position Calibration page on the appears prompting you to select the upper optical
fork position.
2. Click the button corresponding to the upper optical fork position that you set in
Positioning on page
-
click the 850 mm button if the upper optical fork is in the upper position
-
click the 750 mm button if the upper optical fork is in the middle position
-
click the 650 mm button if the upper optical fork is in the lower position
-
click the No Change button if the upper optical fork has not been moved
3. From the Lift Calibration SAVED page that appears, click the OK button.
You are returned to the Calibration: Positioner Summary page.
4. From the Calibration: Positioner Summary page, click the Exit button to leave the Service Desktop.
6-2-5
Check the Lift Lower and Upper Travel Limits
Move the Arm to the lower position, then to the upper position and check whether the Arm stops at the
right positions.
JC-ELE-A-010.fm
Service Information and Procedures Class A 2385072-16-8EN
567, as follows:
Job Card ELE A010 - Set Lift Travel Limits
Page no. 569
569).
Upper Optical Fork
Senographe DS
Chapter 7