GE Healthcare
Revision 1
6-7
Complete a Medical Procedure Card
Perform the following procedure to check for correct installation and configuration of the scanner.
1. Power up the AWS and when prompted at the command line Login: prompt, login as the sdc user as
follows:
a. At the Login: prompt, enter:
sdc
b. At the Password: prompt, press the <Return> key.
2. At the AWS login screen, select clinical from the Username drop-down field, type clinical in the
Password field and click the OK button.
3. Wait until the Browser to appear.
4. Request a new exposure and open the Medical Procedure Card (MPC).
5. Click the first text input field Last Name with the left mouse button to set the keyboard focus.
6. Complete the MPC by scanning all the symbols given on the following page in order, scanning left-to-
right/top-down. As the MPC user interface can change from one AWS release to another, the layout
and entries given in these symbols are templates only: field values are not meaningful and can have
any value. To go to (change focus to) the next input field, scan the $I bar code symbol.
JC-ELE-A-042.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A042 - Bar Code Scanner (Option)
Page no. 723
Senographe DS
Chapter 7