GE Healthcare
Revision 1
14. On the Anode Starter Board 300PL2, remove the cable labeled
connector.
15. Remove the cable labeled
the clamp (15) that secures it to the framework (4 mm allen wrench).
JC-PHY-A-030.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A030 - Generator and Gantry Disconnection Reconnection
XJ2
G300-PL2 XP1 2381701
Page no. 489
G300-PL2 XP1 2381701
15
from the framework of the Generator by releasing
Senographe DS
from the XJ2
Chapter 6