ERR SUB A003 - Gantry Error Messages
Code
Category
102/020
N/A
102/021
HALT
102/022
HALT
102/023
HALT
102/024
HALT
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Text in Log File or
on LCD
STOP LINE WAS
The main CPU detected an error and activated the
ACTIVATED BY
stop motion line or did not deactivated it at all.
CPU
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
LOW VOLTAGE
There is a low voltage error. If the voltage of the
ERROR ON LIFT
power supply falls below 24 V the node sends this
NODE
error.
Change the Lift board then perform Lift Calibration.
LIFT READ TIME-
The MCU has not reacted on the interrupt request
OUT ERROR
(IRQ) coming from the CPLD before 15 ms.
Change the Lift board then perform Lift Calibration.
LIFT CPLD MEM-
There is a CPLD memory error. Test data from the
ORY ERROR
CPLD is periodically read to ensure the proper com-
munication. If this error does not match the node
sends this error.
Change the Lift board then perform Lift Calibration.
UNCONTROLLED
There has been uncontrolled motion. If the motion is
LIFT MOTION
performed when the lift node is not supposed to drive
DETECTED. STOP
it, the lift node sends this error and activates the
LINE ACTIVATED
MCU_STOP_MOTION line which leads to 0x500A
error.
1. Check the lift operation in low speed.
2. If the problem persists, grease the lift screw.
3. If the problem persists, refer to
Possible Cause/Recommended Action
LIFT CONTROL FAILURE on page 948
problem.
Page no. 858
Senographe DS
Job Card
D/R A201
page
1161
CAL A031
page
1687
D/R A201
page
1161
CAL A031
page
1687
D/R A201
page
1161
CAL A031
page
1687
A1013
page 948
JC DIAG A1013
to fix the
ERR-SUB-A-003.fm