Revision 1
Code
Category
112/045
HALT
112/046
DEBUG
ONLY
112/047
DEBUG
ONLY
112/048
DEBUG
ONLY
112/049
DEBUG
ONLY
112/050
DEBUG
ONLY
112/051
DEBUG
ONLY
112/052
DEBUG
ONLY
112/053
DEBUG
ONLY
112/054
DEBUG
ONLY
112/055
DEBUG
ONLY
112/056
DEBUG
ONLY
112/057
DEBUG
ONLY
112/058
DEBUG
ONLY
ERR-SUB-A-003.fm
Text in Log File or
on LCD
ROTATION OPTI-
CAL FORK DEFECT
ROTATION INVALID
USER INTERFACE
ERROR
ROTATION INVALID
LIFT UP SLOW
BUTTON ERROR
ROTATION INVALID
UP FAST BUTTON
ERROR
ROTATION INVALID
LIFT DOWN SLOW
ERROR
ROTATION INVALID
LIFT DOWN FAST
BUTTON ERROR
ROTATION INVALID
ROT LEFT SLOW
BUTTON ERROR
ROTATION INVALID
ROT LEFT FAST
BUTTON ERROR
ROTATION INVALID
ROT RIGHT SLOW
BUTTON ERROR
ROTATION INVALID
ROT RIGHT FAST
BUTTON ERROR
ROTATION INVALID
MLO LEFT BUT-
TON ERROR
ROTATION INVALID
MLO RIGHT BUT-
TON ERROR
ROTATION INVALID
CC BUTTON
ERROR
ROTATION INVALID
COLLIMATOR BUT-
TON ERROR
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
At power-up, the rotation node shuts down the LED
of the left optical fork, and checks that the photore-
sistor shuts down. The optical forks are hardware
limits hence they are never obstructed.
Refer to
JC DIAG A1118 ROTATION CONTROL
FAILURE on page 955
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Page no. 875
ERR SUB A003 - Gantry Error Messages
to fix the problem.