Job Card ELE A042 - Bar Code Scanner (Option)
6-2
Identify and if Necessary Pre-Configure The Bar Code Scanner
Two models of the bar code scanner exist, as follows:
3800LX-14 - has a coiled USB cable
3800G - has an uncoiled (straight) USB cable
The model number (3800LX-14 or 3800G) is located underneath the head of the bar code scanner.
If the model number of the bar code scanner is 3800LX-14, no pre-configuration is required and you
can proceed directly to
If the model number of the bar code scanner is 3800G, the following pre-configuration steps are
required before you can proceed to
1. Ensure your laptop PC is powered off, and connect the bar code scanner with model number 3800G
to a USB connector on your laptop PC.
Note:
Connection or disconnection of the bar code scanner USB cable while the laptop PC is powered
on on can cause damage to the bar code reader and/or the laptop PC.
2. Power on your laptop PC.
3. Hold the scanner handle horizontally directly over the bar code printed below.
4. Press the scanner trigger located below the scanner handle; the scanner red LED beam lights. The
scanner has successfully scanned the symbol if:
-
You hear a beep.
-
The LED beam turns off.
The bar code scanner is now set back to its factory defaults.
5. Hold the scanner handle horizontally directly over the bar code printed below.
6. Press the scanner trigger located below the scanner handle; the scanner red LED beam lights. The
scanner has successfully scanned the symbol if:
-
You hear a beep.
-
The LED beam turns off.
The bar code scanner is now to full ASCII on mode.
7. Power down your laptop PC.
8. Disconnect the bar code scanner from your laptop PC.
Note:
If you want to implement a different code set (for example ISBT 128) or change the bar code mes-
sage length, refer to the bar code scanner documentation.
Chapter 7
Service Information and Procedures Class A 2385072-16-8EN
Connect the Bar Code Scanner on page
Connect the Bar Code Scanner on page
Page no. 716
Senographe DS
720.
720.
JC-ELE-A-042.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Subscribe to Our Youtube Channel

Table of Contents

Save PDF