GE Healthcare
Revision 1
Job Card DIAG A001 - AWS Sun-VTS Diagnostics
4. For V1/V2 Control Stations, plug a
AWS Computer. For V3 Control Stations, plug a
PCI0 on the AWS Computer.
5. Run the full test as follows:
For V1/V2 Control Stations from the command line, enter:
tip /dev/dty/a001s
For V3 Control Stations from the command line, enter:
tip /dev/term/0
The system answers Connected.
6. Type some alphanumeric characters on the keyboard. If the board is good, they are echoed immedi-
ately on the screen
7. To exit the test, type the command:
<Return>~.
Note: It can be necessary to repeat the sequence several times.
8. Reconnect the cable when the test is complete.
12-2 Full Test with Cable
UNIX command
•
Type of test:
1. Power up the system and wait for the console login: prompt on the AWS monitor.
2. Login as root as follows:
a. At the console login: prompt, login as root by entering:
root
b. When requested to specify the root password by entering:
operator
3. Remove the Senographe cable from the Senographe serial port.
4. Plug a
Standard RS-232 Loopback Adapter
5. Run the full test as follows:
For V1/V2 Control Stations from the command line, enter:
tip /dev/dty/a001s
For V3 Control Stations from the command line, enter:
tip /dev/term/0
The system answers Connected.
6. Type some alphanumeric characters on the keyboard. If the cable is good, they are echoed immedi-
ately on the screen
7. To exit the test, type the command:
<Return>~.
Note: It can be necessary to repeat the sequence several times.
8. Reconnect the cable when the test is complete.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Standard RS-232 Loopback Adapter
Standard RS-232 Loopback Adapter
in the end of the cable.
Page no. 984
Senographe DS
in port 1 of Digi Board on the
into Slot #1 of
JC-DIAG-A-001.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.