1. Start up the system in the usual way.
2. From behind the Browser, launch a command window (see
page
120).
3. Open another command window by entering:
xterm &
4. Change to the /export/home/sdc/senovision/logfiles directory where the log files exist, by entering:
cd /export/home/sdc/senovision/logfiles
5. List the log files in the /export/home/sdc/senovision/logfiles directory, by entering:
ls
The filenames of the different log files located in the /export/home/sdc/senovision/logfiles directory
appear.
5-4-2
Compress the Log Files
If the required files appear to be too large to fit on a floppy disk, they can be compressed as follows:
1. Create a temporary directory in the /export/home/sdc/senovision/logfiles directory in which log files
are to be compressed, by entering:
mkdir temp
2. Enter the following commands to copy the log files into the temporary (/export/home/sdc/senovision/
logfiles/temp) directory. The working directory must have been set to logfiles as described above.
Omit commands for log files which do not require compressing:
a. Copy the error.log file by entering:
cp error.log temp
b. Copy the ExamEvents.log file by entering:
cp ExamEvents.log temp
c. Copy the ImageMeasures.log file by entering:
cp ImageMeasures.log temp
d. Copy the SenoServer.log file by entering:
cp trace.SenoServer.log temp
e. Copy the trace.Positioner.log file by entering:
cp trace.Positioner.log temp
f. Copy the PoseidonEvents.log file by entering:
cp PoseidonEvents.log temp
g. If the error.log.back file exists, copy it by entering:
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Launching a Command Line Window on
Page no. 1002
Senographe DS
JC-DIAG-A-005.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.