GE Healthcare
Revision 1
ERR SUB A001 - Generator Error Messages
Code
Message text
079/015
E05 Failure during
exposure
079/017
E16 AWS/SENO
Synchro fail
079/018
I41 Arm error during
exposure
079/019
E05 Failure during
pre-exposure
079/021
E25 AWS/SENO
Synchro. fail
080/001
080/002
080/003
080/004
080/005
080/006
080/007
080/008
080/011
082/000
082/001
082/010
082/020
082/030
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Description
The Generator has not received expo-
sure information from IDC after 15 sec-
onds during an X-ray exposure. This
error occurs only during an X-ray expo-
sure.
The acquisition was not correctly per-
formed on ADS (image not correctly
transferred from IDC, for example)
Can occur during exposure when there
is a problem with the Positioner.
The Generator has not received pre-
exposure information from IDC after 15
seconds.
AWS did not sent acknowledgment to
DMR preparation phase request within
3 seconds.
Gene (IST, exposure): heater fault
Gene (IST, exposure): anode starter
fault
Gene (IST, exposure): power failure
Gene (IST, exposure): interface fault
Gene (IST, exposure): no Gantry reply -
timeout after 1 second
Gene (IST, exposure): Gantry fault
Gene (IST, exposure): no Gantry reply
to exposure command
Gene (IST, exposure): no grid sync
feedback after timeout
The x-ray-authorize RT signal falls dur-
ing an exposure
Gene: 'Endurance' error code
Gene: 'Endurance' error code
Gene: 'Endurance' error code
Gene: 'Endurance' error code
Gene: 'Endurance' error code
Page no. 844
Senographe DS
Recommended actions
Switch Off the system. Switch On
again; repeat the exposure.
Reset system and try again. If the
error persists, check AWS is running
correctly; try rebooting AWS and
Senographe.
On V3 or V4 Control Stations (U20 or
Z400), if the problem persists, addi-
tionally try adding DB9 Adaptors to
the DB9 sockets on the ADS rear
panel (see
Job Card TSG A012 - Use
of DB9 Adaptors to Reduce Intermit-
tent IDC or ADS Communication
Errors on page
1029).
Reboot the Senographe system. If the
problem persists, contact your OLC.
Switch Off the system. Switch On
again; repeat the exposure.
Reboot the Senographe system. If the
problem persists, contact your OLC.
Refer to
Job Card TSG A027 - 080/
011 Error Trouble Shooting on page
1083.
ERR-SUB-A-001.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated