Revision 1
7-2
Monitor Mode Configuration
The 21-inch 3 MP Monitor needs configuring to different settings to the factory defaults, so that:
Landscape mode is enabled (see
only the 1-DICOM mode is available and no other modes can be selected (see
section
7-2-4)
the mode window only appears for 15 seconds (see
the luminosity of the monitor is checked to be 400 cd/m² (see
the control buttons are inactive (to prevent the Operator changing the settings) (see
Note:
If the 21-inch 3 MP Monitor control buttons are locked, you can unlock them according to the pro-
cedure described in
7-2-1
General Controls
This section descibes the general use of the 21-inch 3 MP Monitor control buttons and power LED. The
right hand side of the 21-inch 3 MP Monitor contains various buttons and a LED as shown below.
Existing
Mode
Setting
Item
Function
1
Power Indicator LED
2
Power button
3
Navigational menu buttons
4
Enter button
5
Mode button
Note: The items in bold above are the names of the buttons used in the
procedures below within this Job Card.
JC-ELE-A-008.fm
Service Information and Procedures Class A 2385072-16-8EN
section
7-2-2)
section
7-2-7.
Description
Indicates the monitor's operational
status: Green - ON, Yellow - Power
Saving, not illuminated - OFF.
Turns the monitor power on or off.
Up, Left, Right, and Down buttons
for selecting different monitor configu-
ration menu items and adjustings
A single tap displays the ScreenMan-
ager menu, or confirms the selection
of a menu item.
A double tap saves adjusted values
and closes configuration menu.
Toggles between different preset dis-
play modes
Page no. 553
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
section
7-2-5)
section
Senographe DS
section 7-2-3
7-2-6)
section
Control Panel
and
7-2-7)