GENERATOR CABINET CIRCUIT BREAKERS

The power control buttons on the top of the Generator allow you to power on/off each of the labeled com-
ponents, so that you can isolate them as required. Except for the UPS push-button, the other push-but-
tons contain a second indicator light to the right-hand side of the push-button, and function as follows:
the left-hand side push button toggles on/off the power to/from the component
when power is applied to the component the left-hand side push-button is illuminated green and the
right-hand side light is not illuminated
when power is not applied to the component the right-hand side light is illuminated yellow indicating
that the component is in standby, and the left-hand side push button is not illuminated
The UPS indicator light is off during normal operation when the system is powered up, during which time
it is charging. If the mains AC supply power to the Senographe system is suddenly cut, the UPS supplies
power to the Control Station components for a few minutes. During this time, the UPS indicator light is
illuminated.
Indicators and Switches.fm
Service Information and Procedures Class A 2385072-16-8EN
Conditioner
and MDR
Generator
Page no. 279
MDR Breaker
Conditioner Button
Detector Supply Button
Control Station Button
UPS Button
Generator Breaker
Senographe DS