Revision 1
7-2
Backup Procedure on V3/V4 Control Stations
1. From the Browser, launch the Service Desktop (see
2. From the Service Desktop, click the Utilities button
3. Insert a blank CD-R disk into the writable CD-ROM drive on the AWS unit (not the IDC).
Note:
The presence of a CD-R disk is not essential for the backup. If no CD-R disk is inserted, AWS pa-
rameters are still copied to the hard disk.
4. Select System Tools > Backup. The Back up screen like the following appears.
5. Enter a suitable name into the text field System Backup Description Text; click the START button to
continue. If no CD-R disk is present, the message CD Mount Error Or CD Not Blank appears; click
Continue to continue the backup. The process takes about 10 minutes.
6. An AWS reset is carried out automatically; click OK when it completes.
7. The message System Backup Completed appears at the end of the backup operation.
8. To check that the backup copy is available, click the RESTORE button: The name of the system
backup, Backup 1, appears on the screen.
9. If the backup is not completed successfully, refer to section
10. Remove the CD-R disk; label it with the name and date of the backup and keep it safely.
JC-ELE-A-052.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A052 - Back-up/Restore Procedure
Launching the Service Desktop on page
.
Backup Troubleshooting on page
Page no. 775
Senographe DS
120).
779.
Chapter 7

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF