Job Card PHY A011 - Trackball Option Installation
port (7). Remove the mouse.
10. Remove the mouse mat (9).
11. While holding the handle (10) Remove the three screws (11) securing the mouse mat support (12)
and remove the mouse mat support.
12. Secure the handle (10) with the three new screws (13) that are supplied with the Trackball kit.
13. Install and secure the trackball with the longest screw (14) (length = 38 mm) and spacer (15) that are
supplied with the Trackball Kit.
Chapter 6
Service Information and Procedures Class A 2385072-16-8EN
8
Page no. 446
Senographe DS
9
11
10
12
13
15
10
14
JC-PHY-A-011.fm