GE Healthcare
Revision 1
Code
Category
160/008
HALT
160/009
HALT
160/011
HALT
160/012
HALT
160/013 ADVISORY
161/001
HALT
161/008 ADVISORY
161/009 ADVISORY LCD: A18 Rotation is
161/010 ADVISORY
161/011 ADVISORY
ERR-SUB-A-003.fm
Service Information and Procedures Class A 2385072-16-8EN
Text in Log File or
on LCD
INTERNAL SOFT-
WARE ERROR
INTERNAL SOFT-
WARE ERROR
INTERNAL SOFT-
WARE ERROR
AUTO-REBOOT
MAX RETRY
ATTEMPTS
REACHED
AUTO-REBOOT
RECOVERY
STARTED
INTERNAL SOFT-
WARE ERROR
LCD: A07 Lift is not
operational
not operational
LCD: A03 Compres-
sion is not opera-
tional
LCD: A10 Motion
stopped. Multiple
movements are
requested
Possible Cause/Recommended Action
The Gantry CPU failed to set an attribute.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry CPU failed to publish an event.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
An error code was received from a known node, but
it was not handled by the Gantry CPU in the actual
release.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry CPU failed to boot up on 3 consecutive
attempts.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry CPU failed to boot up and is retrying the
operation.
If system boots successfully, no action is required.
If three reboot attempts are made and fail a 160/012
HALT error is raised.
The Gantry CPU received an unknown UIF request.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry CPU received a lift UIF request while the
lift node is not operational.
Operator error. No service recommended actions.
The Gantry CPU received a rotation UIF request
while the rotation node is not operational.
Operator error. No service recommended actions.
The Gantry CPU received a compression UIF
request while the compression node is not opera-
tional.
Operator error. No service recommended actions.
The compression motion was stopped because the
Gantry CPU received an inconsistent UIF request.
Operator error. No service recommended actions.
Page no. 911
ERR SUB A003 - Gantry Error Messages
Senographe DS
Job Card
Chapter 9
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?