GE Senographe DS Manual page 1089

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
7. Analyze the result:
If [Number of E24 arcing] is smaller than or equals [Accepted Occurrence Number], you shall not
change the tube.
If [Number of E24 arcing] is greater than [Accepted Occurrence Number], you may plan on changing
the tube.
6-2
Acquisition Counter
To display the current acquisition number, you must press the SETUP button.
The current acquisition number appears in the upper right corner of the DMR menu window below
(e.g 2.650E+2, as shown in the illustration below).
JC-TSG-M-028.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG M028 - E24 084/008 Error Troubleshooting
Page no. 1089