GE Healthcare
Revision 1
ERR SUB A003 - Gantry Error Messages
Code
Category
167/009 ADVISORY
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Text in Log File or
on LCD
INCORRECT GAN-
There is currently an incorrect Gantry configuration.
TRY CONFIGURA-
The field of view (FOV) setting does not match the
TION
paddle sliding position, or the paddle/breast support
combination is not correct.
Operator error. No service recommended actions.
Possible Cause/Recommended Action