Job Card ELE A006 - First System Power-On
1
SUPPLIES
None
2
TOOLS
None
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 40 minutes
4
SAFETY PRECAUTIONS
None
5
PREREQUISITES
•
The installed components (AWS, IDS, Conditioner, etc.) must be in a safe state, ready to receive
power, and all of their power switches must be set to on.
•
The instructions in
ensure that the Generator and Mains Distribution Rack power supplies are correctly configured.
6
FIRST POWER ON
After installation of the system, it must be powered up before installing options or configuring the system.
CAUTION
When you are ready to apply power to the Senographe system for the first time, carry out
the following steps:
Note:
If any problem is encountered during the first power up, refer to the checklists for power up/down
given in
Job Card CHK A001 - Boot/Init and Shutdown on page
1. Set the circuit breaker connecting the hospital supply to the Senographe system to on.
2. Set the two circuit breakers in the Generator Cabinet (one in the Generator Rack, the other in the
Main Distribution Rack) to on.
3. To power up the system, press the power switch I on the X-ray Console. Power is applied to the
whole system (Generator, Gantry, Control Station, Digital Detector, etc.). Green power indicators are
lit on the Generator Cabinet, the Control Station, and the LCD Monitor.
JC-ELE-A-006.fm
Job Card ELE A005 - Check Supply Voltages
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A006 - First System Power-On
Page no. 543
Senographe DS
must have been carried out to
1729.
Chapter 7
Chapter 7
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated