ERR SUB A003 - Gantry Error Messages
Code
Category
122/005
HALT
122/006
HALT
122/007
HALT
122/008
HALT
Chapter 9
Text in Log File or
on LCD
COMPR CORRELA-
TION ERROR
COMPRESSION
MOTOR STEP
LOST
COMPRESSION-
OVERSPEED
COMPRESSION
OVERCURRENT
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
The is a Correlation Error. In case signals from abso-
lute coder (potentiometer) and coding wheel (motor
encoder) are not correlated.
See
JC DIAG A1205 COMPRESSION CONTROL
FAILURE on page 957
Occurs when a difference is detected between the
number of steps commanded and the number of
steps detected by absolute encoder. This can hap-
pen whereas correlation is OK, for example when a
mechanical hardpoint imposes zero signal from both
encoder and potentiometer.
See
JC DIAG A1206 COMPRESSION CONTROL
FAILURE on page 958
The compression speed measured by the encoder is
too high in relation to the speed profile (and correla-
tion is OK).
See
JC DIAG A1206 COMPRESSION CONTROL
FAILURE on page 958
Occurs when an integrated value of the current is too
high, that is when the current is too high during too
much time.
See
JC DIAG A1208 COMPRESSION CONTROL
FAILURE on page 960
Page no. 882
to fix the problem.
to fix the problem.
to fix the problem.
to fix the problem.
Senographe DS
Job Card
A1205
page 957
A1206
page 958
A1206
page 958
A1208
page 960