Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7-3-4
DryView 6850 DICOM Printer
•
DICOM printer label: Any name you wish to see appearing on the Film Composer, e.g.,
Carestream 6850.
•
Hostname: Use printer Hostname if any, or a name unique to your Network, e.g., DV6850DRE.
•
Network address: A unique IP address given by the hospital Network Administrator), e.g.,
3.249.15.142.
•
Application Entity Title (AET): Must be DV6850
•
Port number:
•
Layouts:
•
Film size:
•
Pixel size
•
Configuration information Must be LUT=gsdfaa0p.w8u,3
•
Density
•
Magnification type:
•
Smoothing factor:
•
Trim:
•
Printer Memory Size
Note:
To match the new AFSSAPS requirements (in France) for Ambient Light=1 & Illumination=4000,
do not set Review Room into Review Room Management, and check with the printer manufacturer
that the printer is correctly configured.
Chapter 7
Service Information and Procedures Class A 2385072-16-8EN
Must be 5040
Must be 1 x 1
, 1 x 2
Must be 8 in x 10 in
10 in x 12 in
39.08
Must be Min: 20
Must be Cubic
Must be 15
Must be NO
40 (default value)
Page no. 606
, 2 x 1
, or 2 x 2
W = 4924 H = 6224
W = 6224 H = 7526
Max: 360 for DVM film type
Max: 415 for DVM+ film type