GE Senographe DS Manual page 320

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Cable Lay-out and Pin-out
1-84. W504 – Rectangular Blade Optical Sensor Connect
Four cables used to connect the rectangular blade optical sensor connectors (S1 through S4) to the
Interconnection board connectors S1 through S4.
Connectors Sx in the table below correspond to the following:
-
Front blade - S1; Left blade - S2; Rear blade - S3; Right blade - S4:
1-85. W505 – Filter Wheel Optical Sensor Connect
Filter Wheel optical sensor (S5) to Interconnection board S5:
1-86. W506 – Rectangular Blade DC Motor Connect
Four cables, used to connect the rectangular blade DC motor connectors (M1 through M4) to the Inter-
connection board connectors M1 through M4 (board PI 2376693).
Connectors Mx in the table below correspond to the following:
-
Front blade - M1; Left blade - M2; Rear blade - M3; Right blade - M4:
Chapter 3
Service Information and Procedures Class A 2385072-16-8EN
Designation
VCC
LED-K
GND
Zero
Designation
VCC
LED-K
GND
Zero
Designation
Motor -
Motor +
GND
VCC
Channel B
Channel A
Page no. 320
Level
Pin No.
5 V
Sx-1
0/1.3 V
Sx-2
0 V
Sx-3
0/5 V
Sx-4
Level
Pin No.
5 V
S5-1
0/1.3 V
S5-2
0 V
S5-3
0/5 V
S5-4
Level
Pin No.
0/24 V
Mx-1
0/24 V
Mx-2
0 V
Mx-3
5 V
Mx-4
0/5 V
Mx-5
0/5 V
Mx-6
Senographe DS