ERR SUB A003 - Gantry Error Messages
Code
Category
132/028
INHIBIT
132/029
INHIBIT
132/024
DEBUG
ONLY
132/030
DEBUG
ONLY
132/031
DEBUG
ONLY
132/032
DEBUG
ONLY
132/033
DEBUG
ONLY
132/034
DEBUG
ONLY
132/035
DEBUG
ONLY
132/036
DEBUG
ONLY
Chapter 9
Text in Log File or
on LCD
Tilt overcurrent
CPU HEARTBEAT
NOT RECEIVED BY
TILT NODE
HEARTBEAT
DISABLED ON TILT
NODE
TILT INVALID
RX_OPTICAL_FOR
K1_ERROR
TILT INVALID
RX_OPTICAL_FOR
K2_ERROR
TILT INVALID
LIFT_DOWN_PEDA
L_ERROR
TILT INVALID
LIFT_UP_PEDAL_E
RROR
TILT INVALID
COMPRESSION_PE
DAL_ERROR
TILT INVALID
DECOMPRESSION
_PEDAL_ERROR
TILT INVALID
STOP_MOTION_MC
U_ACTIVE ERROR
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
A motion over current is detected.
Refer to
JC DIAG A1305 TILT CONTROL FAILURE
on page 966
and to
Job Card D/R A292 - Tube Tilt
Pivots on page 1397
to fix the problem.
A heartbeat failure occurred. The Gantry CPU did
not send periodic "heartbeat" CAN message.
1. Change the Gantry CPU Board (see
R A190 - Gantry CPU Board on page
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration on
page
1683).
3. Perform a Collimation Check (see
A028 - Collimation Checks on page
The Gantry CPU tried to disable the "heartbeat"
check of the tilt node.
1. Change the Gantry CPU Board (see
R A190 - Gantry CPU Board on page
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration on
page
1683).
3. Perform a Collimation Check (see
A028 - Collimation Checks on page
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Ignore this error. If the problem persists, contact your
OLC.
Page no. 896
Senographe DS
Job Card
A1305
page 966
D/R A292
page
1397
D/R A190
page
1131
Job Card D/
CAL A029
1131).
page
Job Card
1683
ELE A028
Job Card ELE
page 679
679).
D/R A190
page
1131
Job Card D/
CAL A029
1131).
page
Job Card
1683
ELE A028
Job Card ELE
page 679
679).
ERR-SUB-A-003.fm