Monitor Video Signal Checks
JC-TSG-A-001.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A001 - AWS Troubleshooting - No Display on Power Up
From
page 1012
No display on monitor, but monitor
operation LED is lit
Check the video
cable. Is it
defective?
No
Check the monitor
synchronization
cables. Are they
defective or wrongly
connected?
No
Check ADS Computer for
video signal problems
CPU badly seated?
No
RAM badly seated?
No
Defective Graphics
card or Motherboard
video chip?
No
If the problem persists, start again
on
page
1010, or ask for help from
your OLC.
Page no. 1013
Change the video cable.
Yes
Interchange the H sync. and V
sync. cables on the monitor BNC
inputs, or change the cables.
Yes
Reseat the CPU.
Yes
Reseat the RAM.
Yes
Change the ADS Computer
Yes