GE Senographe DS Manual page 347

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
10-2-1. Ancillary Equipment
Consult hospital personnel regarding additional space requirements for hospital equipment such as stor-
age cabinets, sinks, and crash cart.
10-2-2. Positioning of the LCD Monitor on the Control Station
By default, the rotative arm that supports the LCD monitor is installed on the right-hand side of the Con-
trol Station. This is because, for most Operators (who are right-handed), it is more convenient to have
the mouse/trackball on the right-hand side.
However, the following situations require the rotative arm that supports the LCD monitor to be moved on
the left-hand side of the Control Station:
the rare occasion where all or the majority of operators at the customer site are left handed, and the
layout of the room allows the rotative arm to be installed on the left-hand side
the room that hosts the Scenographe system does not permit the Control Station to have the LCD
monitor on the right-hand side, due to either limited space or intrusive objects such as doors or pillars
When reviewing the room that will host the Senographe system, you must determine whether the LCD
monitor must be installed on the left-hand side of the Control Station.
10-2-3. Generic Constraints Overview
The diagram below generically highlights the constraints mentioned above, which you must consider
when planning a room layout.
CU2
5
CU1
6
7
SU2
1. Generator Cabinet.
2. Gantry.
3. Control Station with Radiation Shield.
4. AC power input through power distribution box (supplied by customer).
5. Broadband connection for Insite (supplied by customer).
6. Hospital network connection (supplied by customer).
7. Telephone connection for operators.
After the room layout is decided, suitable provision (plinths, under-floor conduits, etc.) must be made for
passing cables and conduits.
Chap-Pre-Requirements.fm
Service Information and Procedures Class A 2385072-16-8EN
SA1
SA3
3
SE3
Page no. 347
Pre-Installation System Requirements
SE1
SA2
SA2
2
Senographe DS
SU1
SU1
SE2
1
SA1
To door switch
and lights.
Chapter 4
4

Hide quick links:

Advertisement

Table of Contents
loading