GE Senographe DS Manual page 836

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
ERR SUB A001 - Generator Error Messages
Code
Message text
017/004
E03 Rotor failure
017/005
E03 Rotor failure
017/006
E03 Rotor failure
017/007
E03 Rotor failure
017/008
E03 Rotor failure
017/009
E03 Rotor failure
017/010
E03 Rotor failure
017/011
E03 Rotor failure
021/001
021/002
021/003
022/001
E10 Heating failure
track 1
022/002
E10 Heating failure
track 1
022/003
E10 Heating failure
track 1
022/004
E10 Heating failure
track 1
022/005
E10 Heating failure
track 1
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Description
Anode starter: signal Pres_Rotation
(Rotation_Presence) goes 'low' during
rotation at 9000 rpm
Anode starter: signal Pres_Rotation
(Rotation_Presence) goes 'low' during
rotation at 6000 rpm
Anode starter: signal Pres_Rotation
(Rotation_Presence) goes 'low' during
rotation (speed < 9000 rpm)
Anode starter: signal Pres_Rotation
(Rotation_Presence) goes 'low' during
rotation (speed < 6000 rpm)
Anode starter: signal Pres_Rotation
(Rotation_Presence) stays 'high' during
9000 rpm braking
Anode starter: signal Pres_Rotation
(Rotation_Presence) stays 'high' during
6000 rpm braking
Anode starter: signal Pres_Rotation
(Rotation_Presence) stays 'high' during
rotor failure
Anode starter: signal Pres_Rotation
(Rotation_Presence) stays 'low' during
rotation
Heater software error: unknown com-
mand
Heater software error: unknown task
status
Heater software error: Heater scaling
parameters (ICH) or heater values are
not valid
Heater track 1: signal Pres_Chauf
(Heater_Presence) 'high' in standby
status
Heater track 1: signal Pres_Chauf
(Heater_Presence) stays 'low' when the
heater started
Heater track 1: signal Pres_Chauf
(Heater_Presence) goes 'low' during
pre-heater phase
Heater track 1: signal Pres_Chauf
(Heater_Presence) goes 'low' during
first trigger
Heater track 1: signal Pres_Chauf
(Heater_Presence) goes 'low' during
exposure
Page no. 836
Senographe DS
Recommended actions
ERR-SUB-A-001.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF