Job Card TSG A015 - Lift Movement Troubleshooting
2-3
TROUBLESHOOTING STEERING GUIDE FOR TYPE 3 LIFT SCREW ASSEMBLY
Use the following flow charts (illustration 6 and illustration 7) to isolate the issue causing the Lift move-
ment problems on systems with a Type 3 Lift Screw Assembly.
ILLUSTRATION 6 - LIFT TROUBLESHOOTING ENTRY - TYPE 3 LIFT SCREW ASSEMBLY
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Does customer complain about
Error 102/024 or does Error 102/024 repeatly
appear in the error logs?
Yes
Replace Brake Motor
(5413527)
See
Job Card D/R A205 -
Lift Motor on page 1179
Perform Lift Functional Checks.
See
section 3-1
,
Lift Functional
Checks on page
1048. After the
checks, return back here.
NOK
Proceed to the Grease Checks in illustration 7,
and follow the flowchart as required.
Page no. 1046
Perform the lift trouble shooting steps described
No
in illustration 7,
Lift Movement Problems Related to Errors 102/
013 and 102/014 - Type 3 Lift Screw Assembly
on
page 1047
.
OK
OK
Senographe DS
Troubleshooting Flowchart For
JC-TSG-A-015.fm