4. Remove the earth cable (3) from the Generator top cover and leave it to rest on the Conditioner Unit.
Then remove the Generator top cover from the Generator Cabinet.
The front and back covers (wider covers) must be removed before you can remove the left and right
covers.
5. For the front and back covers, do the following:
a. Remove the two bolts (4) that secure the top of the cover (7 mm open ended wrench) and pull it
towards yourself.
JC-PHY-A-042.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A042 - Remove/Reinstall Generator Covers