Text in Log File or
on LCD
Internal 13.3V volt-
age is out of spec
limits
Internal 5V DC volt-
age is out of spec
limits
Negative unregu-
lated voltage is out of
spec limits
Positive unregulated
voltage is out of spec
limits
Internal -5V DC volt-
age is out of spec
limits
Internal 5V DC volt-
age is out of spec
limits
Internal reference
voltage for AD con-
versions is out of
spec limits
Internal 0.6 volts is
bad. Measurement
circuit error
Internal EEPROM
error
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
The internal 13.3V voltage is out of the specified lim-
its.
Ignore this error. If the error persists, contact your
OLC.
The internal 5 V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
The negative unregulated voltage (coming from
transformation) is out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
The positive unregulated voltage (coming from trans-
formation) is out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
The internal -5V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
The internal 5 V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
The internal reference voltage for A/D conversions is
out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
The internal 0.6 V is bad due to a measurement cir-
cuit error.
Ignore this error. If the error persists, contact your
OLC.
The PDU MCU has failed. This error is also gener-
ated when EEPROM is empty. This behavior is nor-
mal during the first 5 s after the PDU is first turned
on.
Ignore this error. If the error persists, contact your
OLC.
Page no. 907
ERR SUB A003 - Gantry Error Messages
Senographe DS
Job Card
Chapter 9
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated